A pilot study exploring the possibility of establishing guidance levels in x-ray directed interventional procedures

Med Phys. 2008 Feb;35(2):673-80. doi: 10.1118/1.2829868.

Abstract

This article summarizes the dosimetric results of an International Atomic Energy Agency coordinated research program to investigate the feasibility of adopting guidance levels for invasive coronary artery procedures. The main study collected clinical data from hospitals located in five countries. A total of 2265 coronary angiograms (CA) and 1844 percutaneous coronary interventions (PCI) were analyzed. Substudies evaluated the dosimetric performance of 14 fluoroscopes, skin dose maps obtained using film, the quality of CA procedures, and the complexity of PCI procedures. Kerma-area product (PKA) guidance levels of 50 and 125 Gy cm2 are suggested for CA and PCI procedures. These levels should be adjusted for the complexity of the procedures performed in a given institution.

MeSH terms

  • Body Burden
  • Feasibility Studies
  • Humans
  • Pilot Projects
  • Practice Guidelines as Topic*
  • Radiation Dosage
  • Radiation Protection / methods*
  • Radiation Protection / standards*
  • Radiography, Interventional / methods*
  • Radiography, Interventional / standards*
  • Radiometry / methods*
  • Radiometry / standards*
  • Relative Biological Effectiveness