Worldwide evaluation of a defibrillation lead with a small geometric electrode surface for high-impedance pacing

Am Heart J. 2003 Dec;146(6):1066-70. doi: 10.1016/S0002-8703(03)00411-3.

Abstract

Background: Pacing leads with a small electrode surface for high-impedance stimulation have been shown to prolong pacemaker longevity, but no sufficient data is available on the safety and feasibility of a defibrillation lead with this novel design.

Methods: We evaluated the clinical performance of a tined, steroid-eluting defibrillation lead with a small electrode surface area (model 6944) in a prospective multicenter study. A total of 542 patients with conventional indications for an implantable cardioverter defibrillator were randomized 1:1 to receive either the model 6944 or a tined, steroid-eluting defibrillation lead with a conventional sized electrode surface area (model 6942). Device performance and electrical parameters were evaluated at implant and 1, 3, 6, and 12 months thereafter (mean follow-up 11.3 +/- 5.6 months).

Results: Baseline characteristics, lead implant success rates, and defibrillation thresholds did not differ significantly between the 2 groups. While pacing thresholds did not differ significantly during follow-up, pacing impedance was approximately twice as high in the model 6944 as in the model 6942 lead (P <.0001). Mean R-wave amplitudes were smaller in patients with a 6944 (9.1 +/- 3.1 mV vs 9.8 +/- 3.6 mV for model 6942, P <.05), but remained stable within both groups throughout the observation period. The total number of ventricular lead-related adverse events and patient survival did not differ significantly between the 2 groups.

Conclusions: The use of a defibrillation lead with a small electrode surface for high-efficiency pacing is safe and feasible and increases pacing impedance without significantly compromising clinical performance.

Publication types

  • Clinical Trial
  • Multicenter Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Canada
  • Defibrillators, Implantable*
  • Electric Impedance
  • Electrodes, Implanted*
  • Equipment Design
  • Equipment Failure
  • Europe
  • Female
  • Heart Diseases / therapy*
  • Humans
  • Male
  • Middle Aged
  • Prospective Studies
  • Time Factors
  • United States