Table 2

Procedural and outcome information

ParameterTotal (n=156)Heart failure (−) (n=144)Heart failure (+) (n=12)P value
Procedural data
 Access route0.067
 Transfemoral133 (85.3)125 (86.8)8 (66.7)
 Transapical20 (12.8)17 (11.8)3 (25.0)
 Transsubcravian1 (0.6)1 (0.7)0 (0.0)
 Direct-Aorta2 (1.3)1 (0.7)1 (8.3)
 Valve type0.753
 Edwards SAPIEN XT46 (29.5)43 (29.9)3 (25.0)
 Edwards SAPIEN 393 (59.6)86 (59.7)7 (58.3)
 Medtronic corevalve3 (1.9)3 (2.1)0 (0.0)
 Medtronic EvolutR14 (9.0)12 (8.3)2 (16.7)
 Valve size (mm)23 (23–26)23 (23–26)23 (23–26)0.604
 Urgency20 (12.8)16 (11.1)4 (33.3)0.027
  Simultaneous catheter measurement
 Mean AVPG before TAVI48 (36–66)48 (35–66)65 (57–96)0.034
 Mean AVPG after TAVI7 (5–11)7 (5–11)8 (6–9)0.795
 Mean AVPG improvement40 (30–59)39 (29–57)60 (49–85)0.023
 LVEDP before TAVI17 (14–23)17 (14–23)21 (15–24)0.474
 LVEDP after TAVI20 (13–25)20 (14–25)16 (10–24)0.529
 PVL grade after TAVI0.162
 None26 (18.6)26 (20.0)0 (0.0)
 Trace79 (56.4)70 (53.8)9 (90.0)
 Mild34 (24.3)33 (25.4)1 (10.0)
 Moderate1 (0.7)1 (0.8)0 (0.0)
 Severe0 (0.0)0 (0.0)0 (0.0)
 Contrast (mL)66 (57–84)66 (56–84)66 (59–74)0.878
 Fluoroscopy time (min)19 (13–29)19 (13–29)21 (12–25)0.939
 Procedure time (min)72 (50–102)71 (50–101)93 (51–129)0.333
 Anaesthesia time (min)132 (109–165)132 (109–162)153 (119–188)0.212
 Volume of infusion (mL)1300 (838–1623)1290 (800–1603)1595 (925–2490)0.204
 Blood transfusion (mL)0 (0–280)0 (0–280)140 (0–615)0.073
 In-out balance (mL)950 (565–1405)940 (550–1355)1395 (785–1940)0.107
Periprocedural complications
 In-hospital death2 (1.3)2 (1.4)0 (0.0)0.681
 Disabling stroke2 (1.3)2 (1.4)0 (0.0)0.681
 Coronary occlusion5 (3.2)5 (3.5)0 (0.0)0.512
 Acute kidney injury8 (5.1)7 (4.9)1 (8.3)0.600
 Permanent pacemaker implantation7 (4.5)6 (4.2)1 (8.3)0.503
  • Caption is the same as in table 1.

  • AVPG, aortic valve pressure gradient; LVEDP, left ventricular end diastolic pressure; PVL, paravalvular leakage; TAVI, transcatheter aortic valve implantation.