Effective orifice area (cm2) | 28.6 | 1.71±0.02 | 1.98±0.02 (115%) | 2.36±0.02 (138%) | 2.03±0.02 (119%) |
50.0 | 2.21±0.02 | 2.24±0.02 (101%) | 2.51±0.03 (114%) | 2.35±0.02 (106%) |
71.4 | 2.35±0.02 | 2.32±0.02 (99%) | 2.55±0.03 (109%) | 2.44±0.02 (104%) |
92.9 (86.2 for TA2) | 2.33±0.02 | 2.42±0.02 (104%) | 2.52±0.03 (108%) | 2.40±0.02 (103%) |
Pressure difference (mm Hg) | 28.6 | 3.3±0.08 | 3.1±0.08 (94%) | 2.1±0.06 (63%) | 2.8±0.07 (85%) |
50.0 | 4.8±0.11 | 4.4±0.10 (92%) | 3.8±0.09 (78%) | 4.3±0.10 (90%) |
71.4 | 8.1±0.14 | 8.2±0.14 (101%) | 7.1±0.12 (88%) | 7.8±0.13 (96%) |
92.9 (86.2 for TA2) | 14.1±0.21 | 11.9±0.18 (84%) | 13.5±0.20 (96%) | 14.7±0.23 (104%) |
Systolic energy losses (mJ) | 28.6 | 5.4±0.3 | 5.3±0.2 (98%) | 7.1±0.3 (131%) | 8.2±0.4 (152%) |
50.0 | 32.8±0.7 | 28.1±0.6 (85%) | 25.8±0.6 (78%) | 31.7±0.7 (97%) |
71.4 | 82.4±2.0 | 78.0±1.8 (95%) | 71.3±1.8 (87%) | 77.8±1.9 (94%) |
92.9 (86.2 for TA2) | 203.8±4.8 | 164.8±4.5 (71%) | 190.5±5.2 (93%) | 227.2±5.8 (111%) |