Treatment of Symptomatic Severe Aortic Stenosis With a Novel Resheathable Supra-Annular Self-Expanding Transcatheter Aortic Valve System

JACC Cardiovasc Interv. 2015 Aug 24;8(10):1359-1367. doi: 10.1016/j.jcin.2015.05.015.

Abstract

Objectives: The purpose of this study was to prospectively evaluate the safety and clinical performance of the CoreValve Evolut R transcatheter aortic valve replacement (TAVR) system (Medtronic, Inc., Minneapolis, Minnesota) in a single-arm, multicenter pivotal study in high- or extreme-risk patients with symptomatic aortic valve stenosis.

Background: Although outcomes following TAVR are improving, challenges still exist. The repositionable 14-F equivalent CoreValve Evolut R TAVR system was developed to mitigate some of these challenges.

Methods: Suitable patients (n = 60) underwent TAVR with a 26- or 29-mm Evolut R valve. Primary safety endpoints were mortality and stroke at 30 days. Primary clinical performance endpoints were device success per the VARC-2 (Valve Academic Research Consortium-2) and the percent of patients with mild or less aortic regurgitation 24 h to 7 days post-procedure.

Results: Patients (66.7% female; mean age 82.8 ± 6.1 years; Society of Thoracic Surgeons Score 7.0 ± 3.7%) underwent TAVR via the transfemoral route in 98.3%, using a 29-mm valve in 68.3% of patients. All attempts at repositioning were successful. No death or stroke was observed up to 30 days. The VARC-2 overall device success rate was 78.6%. Paravalvular regurgitation post TAVR was mild or less in 96.6%, moderate in 3.4%, and severe in 0% at 30 days. Major vascular complications occurred in 8.3%, and permanent pacemaker implantation was required in 11.7% of patients.

Conclusions: The repositionable 14-F equivalent Evolut R TAVR system is safe and effective at treating high-risk symptomatic aortic stenosis patients. Repositioning was successful when required in all patients, with low rates of moderate or severe paravalvular aortic regurgitation and low permanent pacemaker implantation. (The Medtronic CoreValve™ Evolut R™ CE Mark Clinical Study; NCT01876420).

Keywords: aortic valve stenosis; self-expanding heart valve; transcatheter aortic valve replacement.

Publication types

  • Clinical Trial
  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve Insufficiency / etiology
  • Aortic Valve Stenosis / diagnosis
  • Aortic Valve Stenosis / mortality
  • Aortic Valve Stenosis / physiopathology
  • Aortic Valve Stenosis / therapy*
  • Aortic Valve* / physiopathology
  • Arrhythmias, Cardiac / etiology
  • Arrhythmias, Cardiac / therapy
  • Australia
  • Cardiac Catheterization / adverse effects
  • Cardiac Catheterization / instrumentation*
  • Cardiac Catheterization / mortality
  • Cardiac Catheters*
  • Cardiac Pacing, Artificial
  • Female
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / instrumentation*
  • Heart Valve Prosthesis Implantation / mortality
  • Heart Valve Prosthesis*
  • Humans
  • Male
  • New Zealand
  • Prospective Studies
  • Prosthesis Design
  • Risk Factors
  • Severity of Illness Index
  • Stroke / etiology
  • Time Factors
  • Treatment Outcome
  • United Kingdom

Associated data

  • ClinicalTrials.gov/NCT01876420