Biodegradable polymer biolimus-eluting stent versus durable polymer everolimus-eluting stent: a randomized, controlled, noninferiority trial

J Am Coll Cardiol. 2013 Jul 16;62(3):181-190. doi: 10.1016/j.jacc.2013.04.045. Epub 2013 May 15.

Abstract

Objectives: NEXT (NOBORI Biolimus-Eluting Versus XIENCE/PROMUS Everolimus-Eluting Stent Trial) was designed for evaluating the noninferiority of a biolimus-eluting stent (BES) relative to an everolimus-eluting stent (EES) in terms of target lesion revascularization (TLR) at 1 year.

Background: Efficacy and safety data comparing biodegradable polymer BES with durable polymer cobalt-chromium EES are currently limited.

Methods: The NEXT trial is a prospective, multicenter, randomized, open-label, noninferiority trial comparing BES with EES. Between May and October 2011, 3,235 patients were randomly assigned to receive either BES (n = 1,617) or EES (n = 1,618).

Results: At 1 year, the primary efficacy endpoint of TLR occurred in 67 patients (4.2%) in the BES group, and in 66 patients (4.2%) in the EES group, demonstrating noninferiority of BES relative to EES (p for noninferiority <0.0001, and p for superiority = 0.93). Cumulative incidence of definite stent thrombosis was low and similar between the 2 groups (0.25% vs. 0.06%, p = 0.18). An angiographic substudy enrolling 528 patients (BES: n = 263, and EES: n = 265) demonstrated noninferiority of BES relative to EES regarding the primary angiographic endpoint of in-segment late loss (0.03 ± 0.39 mm vs. 0.06 ± 0.45 mm, p for noninferiority <0.0001, and p for superiority = 0.52) at 266 ± 43 days after stent implantation.

Conclusions: One-year clinical and angiographic outcome after BES implantation was noninferior to and not different from that after EES implantation in a mostly stable coronary artery disease population. One-year clinical outcome after both BES and EES use was excellent, with a low rate of TLR and extremely low rate of stent thrombosis.

Trial registration: ClinicalTrials.gov NCT01303640.

Publication types

  • Comparative Study
  • Multicenter Study
  • Randomized Controlled Trial

MeSH terms

  • Absorbable Implants* / standards
  • Aged
  • Aged, 80 and over
  • Coronary Artery Disease / diagnostic imaging
  • Coronary Artery Disease / epidemiology
  • Coronary Artery Disease / surgery*
  • Drug-Eluting Stents* / standards
  • Everolimus
  • Female
  • Follow-Up Studies
  • Humans
  • Male
  • Middle Aged
  • Polymers / administration & dosage*
  • Prospective Studies
  • Radiography
  • Single-Blind Method
  • Sirolimus / administration & dosage
  • Sirolimus / analogs & derivatives*
  • Treatment Outcome

Substances

  • Polymers
  • Everolimus
  • umirolimus
  • Sirolimus

Associated data

  • ClinicalTrials.gov/NCT01303640