The "final" 5-year follow-up from the ENDEAVOR IV trial comparing a zotarolimus-eluting stent with a paclitaxel-eluting stent

JACC Cardiovasc Interv. 2013 Apr;6(4):325-33. doi: 10.1016/j.jcin.2012.12.123. Epub 2013 Mar 20.

Abstract

Objectives: This study sought to report the final 5-year outcomes of the ENDEAVOR IV (A Randomized, Controlled Trial of the Medtronic Endeavor Drug [ABT-578] Eluting Coronary Stent System Versus the Taxus Paclitaxel-Eluting Coronary Stent System in De Novo Native Coronary Artery Lesions) trial comparing the Endeavor zotarolimus-eluting stent (E-ZES) (Medtronic, Santa Rosa, California) with the Taxus paclitaxel-eluting stent (PES) (Boston Scientific, Natick, Massachusetts) in patients with single de novo coronary lesions.

Background: Primary results of the ENDEAVOR IV trial demonstrated similar clinical outcomes with E-ZES and PES. Concerns with regard to late adverse clinical events with drug-eluting stents highlight the need for long-term follow-up with these devices.

Methods: Late outcomes after the use of E-ZES and PES were examined in the multicenter randomized ENDEAVOR IV trial in cumulative and landmark analyses. Assessed outcomes were related to device efficacy and patient safety.

Results: At 5 years, clinical data were available for 722 (93.4%) E-ZES patients and 718 (92.6%) PES patients. Overall rates of target lesion revascularization (7.7% vs. 8.6%, p = 0.70) and target vessel failure were similar (17.2% vs. 21.1%, p = 0.061) with E-ZES compared with PES. The incidence of cardiac death or myocardial infarction (MI) was lower with E-ZES (6.4% vs. 9.1%, p = 0.048), primarily driven by a lower rate of target vessel MI with E-ZES (2.6% vs. 6.0%, p = 0.002). Although overall definite/probable stent thrombosis rates were similar between stents (1.3% vs. 2%, p = 0.42), rates of very late stent thrombosis (0.4% vs. 1.8%, p = 0.012) and late MI events (1.3% vs. 3.5%, p = 0.008) were significantly lower with E-ZES compared with PES.

Conclusions: These data demonstrate the durable efficacy and safety of E-ZES compared with PES for the treatment of de novo coronary lesions. Significant improvements in late safety outcomes were observed with E-ZES but should be considered hypothesis-generating, given the limited statistical power of the trial. (The ENDEAVOR IV Clinical Trial: A Trial of a Coronary Stent System in Coronary Artery Lesions; NCT00217269).

Publication types

  • Comparative Study
  • Multicenter Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Cardiovascular Agents / administration & dosage*
  • Coronary Artery Disease / mortality
  • Coronary Artery Disease / therapy*
  • Coronary Thrombosis / etiology
  • Coronary Thrombosis / mortality
  • Drug-Eluting Stents*
  • Female
  • Humans
  • Kaplan-Meier Estimate
  • Male
  • Middle Aged
  • Myocardial Infarction / etiology
  • Myocardial Infarction / mortality
  • Paclitaxel / administration & dosage*
  • Percutaneous Coronary Intervention / adverse effects
  • Percutaneous Coronary Intervention / instrumentation*
  • Percutaneous Coronary Intervention / mortality
  • Proportional Hazards Models
  • Prospective Studies
  • Prosthesis Design
  • Risk Factors
  • Single-Blind Method
  • Sirolimus / administration & dosage
  • Sirolimus / analogs & derivatives*
  • Time Factors
  • Treatment Outcome
  • United States

Substances

  • Cardiovascular Agents
  • zotarolimus
  • Paclitaxel
  • Sirolimus

Associated data

  • ClinicalTrials.gov/NCT00217269