Optical coherence tomography versus intravascular ultrasound to evaluate coronary artery disease and percutaneous coronary intervention

JACC Cardiovasc Interv. 2013 Mar;6(3):228-36. doi: 10.1016/j.jcin.2012.09.017.

Abstract

Objectives: We compared intravascular ultrasound (IVUS) and 2 different generations of optical coherence tomography (OCT)-time-domain OCT (TD-OCT) and frequency-domain OCT (FD-OCT)-for the assessment of coronary disease and percutaneous coronary intervention (PCI) using stents.

Background: OCT is a promising light-based intravascular imaging modality with higher resolution than IVUS. However, the paucity of data on OCT image quantification has limited its application in clinical practice.

Methods: A total of 227 matched OCT and IVUS pull backs were studied. One hundred FD-OCT and IVUS pull backs in nonstented (n = 56) and stented (n = 44) vessels were compared. Additionally, 127 matched TD-OCT and IVUS images were compared in stented vessels.

Results: FD-OCT depicted more severe native coronary disease than IVUS; minimal lumen area (MLA) was 2.33 ± 1.56 mm(2) versus 3.32 ± 1.92 mm(2), respectively (p < 0.001). Reference vessel dimensions were equivalent between FD-OCT and IVUS in both native and stented coronaries, but TD-OCT detected smaller reference lumen size compared with IVUS. Immediately post-PCI, in-stent MLAs were similar between FD-OCT and IVUS, but at follow-up, both FD-OCT and TD-OCT detected smaller MLAs than did IVUS, likely due to better detection of neointimal hyperplasia (NIH). Post-PCI malapposition and tissue prolapse were more frequently identified by FD-OCT.

Conclusions: FD-OCT generates similar reference lumen dimensions but higher degrees of disease severity and NIH, as well as better detection of malapposition and tissue prolapse compared with IVUS. First-generation TD-OCT was associated with smaller reference vessel dimensions compared with IVUS.

Publication types

  • Comparative Study

MeSH terms

  • Aged
  • Analysis of Variance
  • Cell Proliferation
  • Coronary Restenosis / diagnostic imaging
  • Coronary Restenosis / etiology
  • Coronary Restenosis / pathology
  • Coronary Stenosis / diagnosis*
  • Coronary Stenosis / diagnostic imaging
  • Coronary Stenosis / pathology
  • Coronary Stenosis / therapy*
  • Coronary Vessels / diagnostic imaging*
  • Coronary Vessels / pathology*
  • Female
  • Humans
  • Hyperplasia
  • Linear Models
  • Male
  • Middle Aged
  • Neointima
  • Percutaneous Coronary Intervention* / adverse effects
  • Percutaneous Coronary Intervention* / instrumentation
  • Predictive Value of Tests
  • Severity of Illness Index
  • Stents
  • Time Factors
  • Tomography, Optical Coherence*
  • Treatment Outcome
  • Ultrasonography, Interventional*