Elsevier

Journal of Vascular Surgery

Volume 54, Issue 3, September 2011, Pages 685-690
Journal of Vascular Surgery

Clinical research study
From the Society for Vascular Surgery
Prevalence and clinical significance of stent fracture and deformation following carotid artery stenting

Presented at the 2009 Vascular Annual Meeting of the Society for Vascular Surgery, Denver, Colo, June 11-14, 2009.
https://doi.org/10.1016/j.jvs.2011.03.257Get rights and content
Under an Elsevier user license
open archive

Objective

Carotid artery stenting (CAS) is a developing intervention for carotid artery stenosis, and long-term outcomes remain unclear. We examined the prevalence and clinical significance of carotid stent fractures or deformations following CAS.

Methods

Two hundred thirty-one CAS performed in 219 patients at one academic institution between August 2000 and March 2009 were reviewed. One hundred sixteen stents (57 closed cell, 59 open cell) were evaluated with multiplanar plain films of the neck to assess for stent fracture or deformation. Stent fracture was identified by wire strut disruption. Deformation was defined as an increase in stent-cell area from stent strut distortion. Study endpoints included rate of stent fracture or deformation determined using Kaplan–Meier and life table analysis. Factors associated with stent fracture or deformation were identified by Cox regression. Effect of stent fracture or deformation on recurrent carotid artery stenosis >70% requiring reintervention and postoperative stroke was studied.

Results

There were five stent fractures (4%) and 27 deformed stents (23%). Rate of stent fracture or deformation was 15% at 2 years and 50% at 4 years. Deformations were significantly more common in open cell stents than in closed cell stents (58% vs 5% at 4 years, P < .00005). Presence of calcified plaque on plain film was significantly associated with increased rate of stent fracture or deformation (P = .0006). At median follow-up of 25 months, restenosis requiring treatment occurred in four (5%) and late stroke in one (1%). Neither stent fracture nor deformation was associated with late stroke or reintervention.

Conclusions

Stent fracture and deformation is not uncommon following CAS and is associated with the presence of heavy calcification. Whether a carotid stent fractures or deforms correlates with stent design. Larger studies are necessary to determine the possible clinical significance of carotid stent fracture and deformation.

Cited by (0)

Competition of interest: Dr Powell is a consultant for Anges.

The editors and reviewers of this article have no relevant financial relationships to disclose per the JVS policy that requires reviewers to decline review of any manuscript for which they may have a competition of interest.