Table 4

Effect of postdilation on stent dimensions assessed by FD-OCT and IVUS (N=6)

IVUSFD-OCT
PrePostp ValuePrePostp Value
In-stent diameter, mm
 Minimum2.55±0.422.70±0.500.0572.60±0.412.82±0.340.036
 Maximum4.43±0.634.77±0.850.1523.84±0.484.25±0.650.031
In-stent area, mm2
 Minimum7.73±2.207.95±2.720.5517.30±1.628.15±1.900.023
 Maximum13.02±3.4114.77±5.450.18410.72±2.5513.00±3.620.040
Percentage of stent malapposed (%)7±77±40.98434±1219±100.004
  • FD-OCT, frequency domain optical coherence tomography; IVUS, intravascular ultrasound.