Long and short of optimal stent design
  1. Timothy Watson1,2,
  2. Mark W I Webster2,3,
  3. John A Ormiston2,3,
  4. Peter N Ruygrok2,3 and
  5. James T Stewart2,3
  1. 1Department of Cardiology, Tan Tock Seng Hospital, Singapore, Singapore
  2. 2Faculty of Medical and Health Sciences, University of Auckland, Auckland, New Zealand
  3. 3Green Lane Cardiovascular Service, Auckland City Hospital, Auckland, New Zealand
  1. Correspondence to Dr Timothy Watson; timothy_james_watson{at}ttsh.com.sg

Abstract

The ideal stent must fulfil a broad range of technical requirements. Stents must be securely crimped onto the delivery balloon and, in this form, must have a low profile and be sufficiently flexible to facilitate deliverability to the lesion site without distortion or displacement. Following expansion, stents must exert sufficient radial force on the vessel wall to overcome lesion resistance and elastic recoil. To achieve an optimal lumen diameter, the lesion must be uniformly and adequately scaffolded, with minimal tissue prolapse between struts but without compromising side-branch access. Furthermore, the deployed stent must conform to the vessel curvature to minimise vessel distortion, particularly at the stent edges. Radio-opacity is also important to guide safe positioning, adequate deployment and postdilataion and to permit assessment of optimal stent expansion. Equally though, the stent lumen must also be sufficiently visible to allow radiographic assessment of flow dynamics and restenosis. Efforts to optimise one characteristic of stent design may have detrimental effects on another. Thus, currently available stents all reflect a compromise between competing desirable features and have subtle differences in their performance characteristics. Striving to achieve stents with optimal deliverability, conformability and radial strength led to a reduction in longitudinal strength. The importance of this parameter was highlighted by complications occurring in the real-world setting where percutaneous coronary intervention is often undertaken in challenging anatomy. This review focuses on aspects of stent design relevant to longitudinal strength.

  • percutaneous coronary interventions
  • drug eluting stents
  • longitudinal deformation
  • bench testing

This is an Open Access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Footnotes

  • Competing interests None declared.

  • Provenance and peer review Commissioned; internally peer reviewed.

  • Data sharing statement There are no additional data available for this paper.

Linked Articles