Abstract
Objectives Inappropriate shocks are unpleasant and painful. We hypothesise that remote monitoring and careful attention to known and incident atrial fibrillation (AF) can reduce inappropriate shocks to a very low level in clinical praxis.
Methods Altogether 259 patients with implantable cardioverter defibrillator implanted for secondary (S, n=113) and primary (P, n=146) prevention were followed via remote monitoring. At implant, 42S (37%) and 54P (37%) patients had known AF.
Results Inappropriate shocks, all but five due to AF, occurred in 7S (6.2%) and 11P (7.5%), and there were only inappropriate shocks in 5/7S and in 8/11P. They occurred in four of 42S (9.5%) with and in three of 71S (4.2%) without known AF, and in seven of 54P (13%) with and in four of 92P (4.3%) without known AF. The median time from shock to action was 5 and 1 day, respectively. Actions were medication with amiodarone, β blockers, β blockers+amiodarone or β blockers+digoxin (n=5), β blockers+insertion of an atrial lead (n=1), replacement of a fractured lead (n=2), reprogramming in combination with β blockers, digoxin or amiodarone (n=4), reprogramming (n=2) and none (n=4). After action, four further inappropriate shocks occurred during more than 2 years of follow-up, all due to AF.
Conclusions Inappropriate shocks occurred at a low rate and most often because of AF known at implant. Remote monitoring enabled rapid action, after which few inappropriate shocks occurred over more than 2 years. Attention to known and incident AF was the most important action to reduce inappropriate shocks.
This is an Open Access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/